REXEED-L


Gamma sterilized wet-type dialyzer
 
Best biocompatible membrane

Clearance(in vitro)

 

  REXEED-13L REXEED-15L
Blood Flow (mL/min) 200 300 400 200 300 400
Clearances (mL/min)1) Urea 192 255 291 195 265 305
Creatinine 186 241 276 191 251 287
Phosphate 147 175 192 157 191 205
Vitamin B12 88 98 100 97 108 113
KoA (Urea)(mL/min)2) 888 1045
KUF (mL/hr/mmHg (mL//hr/kPa))3) 10 (75) 12 (90)
Effectives Surface Area (m2) 1.3 1.5
  REXEED-18L REXEED-21L
Blood Flow (mL/min) 200 300 400 200 300 400
Clearances (mL/min)1) Urea 197 270 312 198 277 325
Creatinine 192 256 297 195 268 314
Phosphate 162 199 215 171 213 236
Vitamin B12 103 117 122 115 128 136
KoA (Urea)(mL/min)2) 1145 1321
KUF (mL/hr/mmHg (mL//hr/kPa))3) 15 (113) 18 (135)
Effectives Surface Area (m2) 1.8 2.1

Conditions:
1) Clearances: Qd=500mL/min, Qf=0mL/min
2) KoA: Qb=300mL/min, Qd=500mL/min, Qf=0mL/min
3) KUF: bovine blood, TP=60±5g/L, Hct=32±2%, Qb=300mL/min
ISO 8637:2004

Lineup

REXEED-13L
REXEED-18L

REXEED-15L
REXEED-21L

 

 

REXEED-L specifications

  REXEED-13L REXEED-15L
Membrane REXBRANE™ (Asahi Polysulfone)
Internal Diameter of Hollow Fiber (μm) 185
Wall Thickness of Hollow Fiber (μm) 45
Priming Volume (mL) 76 86
Maximum TMP (mmHg (kPa)) 500 (66)
Maximum Blood Flow (mL/min) 500
Maximum Dialysate Flow (mL/min) 800
Dimensions (mm[L] x mm[D]) 334 x 38 334 x 41
Weight (g) 412 463
Sterilization Gamma-Ray
  REXEED-18L REXEED-21L
Membrane REXBRANE™ (Asahi Polysulfone)
Internal Diameter of Hollow Fiber (μm) 185
Wall Thickness of Hollow Fiber (μm) 45
Priming Volume (mL) 103 117
Maximum TMP (mmHg (kPa)) 500 (66)
Maximum Blood Flow (mL/min) 500
Maximum Dialysate Flow (mL/min) 800
Dimensions (mm[L] x mm[D]) 334 x 43 334 x 47
Weight (g) 520 604
Sterilization Gamma-Ray

Note: High permeability devices. Use only with ultrafiltration controlling equipment. ISO 8637:2004

James B.Post, MD.
Thrombocytopenia Associated With Use of a Biocompatible Hemodialysis Membrane: A Case Report.
American Journal of Kidney Diseases, Vol 55, No 6 (June), e25-e28, 2010
Hoshiyama et al.
A study of dialysis conditions in patients with high body weight.
Jpn. J. High Performance Membrane '05, Kidney and Dialysis, Suppl., 59:39-43, 2005
Abe et al.
Clinical evaluation of plasma insulin and C-peptide levels with 3 different high-flux dialyzers in diabetic patients on hemodialysis.
The international JOURNAL OF ARTIFICIAL ORGANS Vol.31, No. 10, pp. 898-904, 2008